Critical Product Correction Notice

HeartSine Samaritan PAD Pad-Pak

22 September, 2025

Stryker has initiated a Critical Product Correction for the HeartSine Pad-Pak (battery and electrode pack) used in HeartSine samaritan PADs (HeartSine AED).

This applies to Pad-Paks with expiry dates between 17 April 2025 – 01 August 2029. Customers are being asked to check for bent locator pins and to confirm the Pad-Pak is correctly inserted into the AED.

Issue

Some Pad-Paks may:

  • Not be correctly inserted into the AED, or
  • Have bent locator pins

 

This could result in:

  • Device not powering on
  • Device repeatedly prompting “apply pads to patient’s bare chest
  • Failure to deliver therapy in an emergency
 
Background
  • The HeartSine Pad-Pak is a combined battery + electrode cartridge.
  • Since 2018, 36 global events have been reported (5 confirmed bent pins, others linked to user error).
 
Action for Customers
  • Inspect all Pad-Paks within the affected expiry range.
  • Check for bent locator pins and confirm correct insertion.
 
Action for Sales Team
  • Reassure customers this is a precautionary correction.
  • Direct customers to the official Stryker communication to determine if their device is effected and() for detailed instructions. View Lookup Form HERE.
  • Escalate any issues or confirmed faulty units to: